Customer needs

A highly pure copper material containing no oxide, such as oxygen free copper (OFC) or tough pitch copper (TPC), which is 99.95% or higher, is suitable for high-end acoustic equipment, signal cable, etc.  Oxygen in copper affects parts performance and quality. Hence, accurate management of lower oxygen concentration is essential. Is there a way to quickly and accurately quantify oxygen in ppm order?

Raw Copper Material

Solved by EMGA series !

Oxygen to trace amount ranges in conformity with JIS H 1067.

Electric wire

Result Sample: copper

timeSample weight (g)Oxygen (mass ppm)

1

1.0194.7

2

1.0234.8

3

1.0225.2

4

1.0165.1
51.0125.1
Average5.0

Standard deviation

0.2

Variation coefficient (%)

4.4
Range0.5
scrollable

Customer feedback

By using EMGA, it is possible to analyze the low concentration of oxygen in minutes.