Dissolved Oxygen Monitor in Low Concentration HF HD-960L

HF(氫氟酸)DO(溶氧)監測儀/純水DO(溶氧)監測儀

HD-960L

這是一款專為低流量應用而設計的溶解氧監測儀,非常適合在濕式製程中監測 HF 溶液中的溶氧濃度。

配備量程切換功能,始終以最佳量程/分辨率測量從低到高的溶解氧濃度,使其能夠適應每個製程的特定測量要求。

DO

DC24VTransmission OutputContact OutputRS-485Single Component

事業部: Semiconductor
產品分類: Wet Process Control
製造商: HORIBA Advanced Techno Co., Ltd.

Automatic Range Switching

Automatically switches between three measurement ranges, from μg/L order to mg/L order, enabling optimal resolution measurements.

Suitable for Measuring Small Sample Volumes

Can measure flow rates as low as 15 mL/min.

Simple Calibration Function

Equipped with zero calibration for electricity and zero/span calibration functions for gas.

Model NameHD-960L
Measurement range RepeatabilityMeasurement rangeRepeatability
0.0 to 200.0 μg/L+/-1% FS
200 to 2000μg/L+/-1% FS
2.00 to 20.00mg/L+/-1% FS
Conditions measurement sampleSample Pressure0 to 0.1MPa
Sample flow rate15 to 200ml /min.
Sample temperature10 to 45 °C
Sample HF concentration0 to 5000ppm
Transmission output4-20mA, 4points
Contact output4 points
Power supplyDC24V +/- 10% 15W
Combined Sensor unitDO-100
Combined Sensor#5600
Sample inlet/outlet1/4 inch tube
scrollable

Schematics

留言諮詢

如您有任何疑問,请在此留下詳細需求或問題,我們將竭誠您服務。

* 這些欄位為必填項目。