APNA-370/CU-2

Ambient Ammonia Monitor

APNA-370/CU-2 Ambient Ammonia Monitor measures increased nitrogen oxide (NOx) raected through oxide catalyst treatment as Ammonia (NH3) concentration with chemiluminescent (CLD) method. HORIBA's original Cross-flow modulation method ensures long-term stability and reliable measuring value with 5ppb for min. detectable sensitivity.

 

Division: Ambient
Manufacturing Company: HORIBA, Ltd.
Base product: AP-370 Series
  • Contiuous monitoring for Ammonia in ambient air with high sensitivity.

 

ModelAnalyzer unit: APNA-370
Converter unit: CU-2
PrincipleOxidationcatalyst + Chemiluminescence
Range0-(0.1/0.2/0.5)/1 ppm or 0-1/2/5/10 ppm
Min. detectable sensitivity5 ppb
Repeatability±2.0%of F.S.
Linearity±2.0% of F.S.
Zero drift±2.0% of F.S./day
Span drift±2.0% of F.S./day
Sample gas flow rateApprox. 3.0L/min
IndicationMeasuring value, alarm
AlarmsAPNA-370: Calibration error, Battery error, Flow rate error, Pressure error, Converter temperature error, Power failure
CU-2: Catalyzer temperature error
Input/output0-1V, (2 lines from instant value, and integrated value or average value)
Contact inputoutput (range, mode, external reset, telemeter failure, alarm, etc.)
RS-232C (option)
Power100/110/115/120/220/230/240 VAC, 50/60 Hz
APNA-370: approx. 200VA
CU-2: approx. 200VA
DimensionsAPNA-370: 430(W) x 550(D) x 221(H) mm
CU-2: 430(W)x450(D)x310(H) mm
MassAPNA-370: approx. 21kg
CU-2: approx. 20kg
scrollable
Monitoring of Airborne Molecular Contamination in Clean Room of Electronics Industry
Monitoring of Airborne Molecular Contamination in Clean Room of Electronics Industry
Monitoring of Ammonia Airborne Molecular Contamination in Semiconductor Fabrication
Monitoring of Ammonia Airborne Molecular Contamination in Semiconductor Fabrication

제품 문의

HORIBA제품의 자세한 정보를 원하시면, 아래의 양식에 내용을 입력을 부탁드립니다.

* 는 필수입력항목입니다.

Related products

FTX-ONE-CL
FTX-ONE-CL

FTIR 배기 가스 분석기 희석 측정 타입

FTX-ONE-CS/RS
FTX-ONE-CS/RS

FTIR 배기가스 분석 측정 시스템 (직접 측정 방식)

IR-200
IR-200

Gas Monitor for Chamber Cleaning End Point Monitoring

IR-400
IR-400

High-grade type Gas Monitor for Chamber Cleaning End Point Monitoring

MEXA-ONE
MEXA-ONE

배기 가스 측정 장치

MEXA-ONE-QL-NX
MEXA-ONE-QL-NX

레이저 분광 배기가스 측정 장치

OBS-ONE GS Unit
OBS-ONE GS Unit

차량 탑재형 배기가스 측정 장치

OBS-ONE IRLAM Unit
OBS-ONE IRLAM Unit

NH3/N2O 대응용 차량 탑재형 배기가스 측정 장치

Show More

ENDA-7000 Series
ENDA-7000 Series

굴뚝 배출가스 분석장치

MEXA-ONE
MEXA-ONE

배기 가스 측정 장치

OBS-ONE GS Unit
OBS-ONE GS Unit

차량 탑재형 배기가스 측정 장치

OBS-ONE IRLAM Unit
OBS-ONE IRLAM Unit

NH3/N2O 대응용 차량 탑재형 배기가스 측정 장치

Show More

Corporate