Optical Performance Specifications | |
---|---|
Optical power | > 300 m W (grating, bandpass & wavelength dependent) |
Spot size at slit exit | 10 mm (slit dependent) |
Diverging beam angle (full) | 14.4 degrees |
Numerical aperture (N.A.) | 0.12 |
Optical noise | From 0.15% to 0.2% RMS |
Optical stability | 2% |
Other Specifications | |
Input | 210–240 V AC 50/60 Hz |
Starting | 45 kV starting pulse |
Power rating | 800–1200 watts (adjustable) — recommended 800–1000 watts |
Lamp module type | 1000 W Xenon, 1000 W Mercury/Xenon (proprietary to HORIBA) |
Lamp life | Typically 1,500 hrs |
Focusing optics | High efficiency f/4 ellipsoid reflector |
Power precision | 0.04% (0.4 watts) |
Output volts compliance | 17–23 VDC |
Output current limit | 70 A rms |
Dimensions | 489 x 375 x 329 mm (19.3 x 14.8 x 12.9 inches) |
Weight | 31 kg (68 pounds) |
Window diameter (D) | 127 mm (5.0 inches) |
Center beam line height (without feet) | 128 mm (5.0 inches) |
Optical Motor Controller Specifications | |
Power supply | Universal Power Supply included |
TTL output | Synchronization TTL output each time motor stops |
Stepper motor | Two phase motor 1 A per phase, 200 steps per revolution, 1.8 degrees per step |
Maximum motor speed | 1200 RPM with zero torque |
Maximum speed with mono | 15,000 nm per minute |
Stepping motor voltage | 5–12 V |
Stepping modes | Full, Half and Micro steps: 1/8, 1/16, 1/32, 1/64 computer selectable |
Slew rate | 1 to 62,500 micro steps per second |
Calibration | Auto calibration of wavelength |
Ramping | Linear ramping rate for heavy duty, fast, precision operation |