DeltaFlex

TCSPC/MCS Fluorescence Lifetime System

Ultimate TCSPC/MCS Performance and Flexibility for Lifetimes From 5 Picoseconds to Seconds

The HORIBA DeltaFlex™ system is the Time-Correlated Single Photon Counting (TCSPC) system of choice for the measurement of fluorescence lifetimes from picoseconds to seconds. Taking advantage of more than 45 years of TCSPC innovation, it is a complete fluorescence/phosphorescence system designed and optimized with innovative pulsed lasers and LEDs, timing electronics, detectors and sample handling components, all powered by HORIBA’s highly intuitive and extremely powerful EzTime™ touchscreen software interface. 

Segment: Scientific
Manufacturing Company: HORIBA Scientific

Unique Benefits from a Lifetime in Fluorescence

  • Flexibility to meet any requirement
  • Better user experience with automated component recognition and control
  • Extensive choice of sources and detectors to meet all wavelength and lifetime requirements
  • EzTime intuitive, automated, touchscreen software


DeltaFlex: Flexibility is in the name, and flexibility is in the design

The modular DeltaFlex system is comprised of a choice of the following main components:

  • Optical configuration
    Choose from filters or monochromators and single detector (L format) or simultaneous dual detector (T Format)
  • Excitation sources
    Choose from a plethora of pulsed lasers and LED sources
  • Detection modules
    Choose from a wide range of dedicated TCSPC detectors depending on your emission wavelength range, lifetime requirements and budget
  • Timing electronics
    Choose from standard or high resolution electronic interfaces depending on your lifetime requirements

 

All DeltaFlex systems are controlled using EzTime software which provides for full instrumentation control and acquisition, as well as a comprehensive suite of data analysis modules for fluorescence and phosphorescence lifetime determination, decay associated spectra (DAS), time-resolved anisotropy and uncorrected steady state emission spectra (if equipped with a scanning emission monochromator).

Optional Pulsed Laser and LED Light Sources

 

Featured Videos

 

DeltaDiode™ General Specifications

DeltaDiode TypePulse DurationRepetition RateAvailable Wavelengths
DeltaDiode Laser
(Denoted with an “L” on part number)
35 to 200 ps10 kHz to 100 MHz375 to 1,310 nm
DeltaDiode LED
(No “L” on part number)
750 to 950 ps10 kHz to 25 MHz265 to 455 nm

 

DeltaDiode Controller Specifications

DD-C1 Controller

FunctionSpecification

Repetition Rates

10 kHz, 20 kHz, 50 kHz, 100 kHz, 250 kHz, 500 kHz, 1 MHz, 2 MHz, 4 MHz, 5 MHz, 8 MHz, 10 MHz, 16 MHz, 20 MHz, 25 MHz, 50 MHz, 80 MHz, 100 MHz, or Trigger Input (single-shot to 50 MHz). Subject to attached head

Trigger Input

Pulse amplitude +0.5 V to +5 V, trigger threshold software programmable from +0.2 V to +2 V, 50 Ω, 20 ns minimum spacing

Sync Outputs

Simultaneous output of NIM-compatible (-0.8 Vpp 50 Ω) and TTL-compatible (+2 Vpp 50 Ω), automatic width selection 4-15 ns nominal

Sync Delay Control

Adjustment of sync output pulse timing in range -10 ns to +10 ns nominal in 1 ns steps, uncalibrated

Fast Gate Input

Pulse amplitude +0.5 V to +5 V, trigger threshold software programmable from +0.2 V to +2 V, 50 Ω. Selectable Inhibit/Enable modes

Slow Gate Input

Pulse amplitude +2 V to +5 V. 10 kΩ. Selectable Inhibit/enable modes. Operates in Pulsed and CW modes

Interlock

2-pin connector (included). Contacts must be short-circuited to enable emission

Connection to Head

1.5 m cable (included)

User Interface

LCD display (stand-alone operation) or software (PC control)

PC interface

USB 2.0 with integral hub for downstream connection to other USB peripherals (cable to host PC and software supplied)

Power Requirement

90 V to 250 V AC, 50/60 Hz, 100 VA

Operating Temperature

+15˚ C to +30˚ C (ambient)

Weight & Dimensions

3.1 kg, 234 x 255 x 92 mm

 

Solas MOFA Fiber Laser Specifications

Parameter

Solas 355L

Solas 532L

Solas 1064L

Power

> 7 mW at 80 MHz

> 30 mW at 80 MHz

> 300 mW at 80 MHz

Power Stability

0.5% RMS over 10 min

Pulse duration

80 to 100 ps (FWHM)

Spectral Width

< 0.2 nm

< 0.4 nm

Resolution Rate

Pulse on demand to 100 MHz

Pointing Stability

< 1 μrad

Polarization Extinction Ratio

> 20 dB

Beam Quality

Multi-mode fiber output

Circular TEM00 beam

 

SpectraLED Specifications

 Pulse DurationRepetition RateAvailable Wavelengths
SpectraLED< 250 ns to > 10 s0.1 to 100 kHz265 to 1,275 nm

 

SpectraXE Specifications

 Pulse DurationRepetition RateAvailable Wavelengths
SpectraXE0.4 μs0.1 Hz to 80 Hz185 to 2,000 nm

 

HPPD Specifications

ModelWavelength RangeTemporal Response
(IRF FWHM)
Dark CountQuantum Efficiency
HPPD-650220 to 650 nm50 ps< 100 cps28% (340 nm)
HPPD-720300 to 720 nm120 ps< 1000 cps47% (530 nm)
HPPD-860220 to 860 nm50 ps< 200 cps23% (280 nm)
HPPD-870300 to 870 nm130 ps< 500 cps26% (630 nm)
HPPD-890380 to 890 nm160 ps< 1000 cps16% (630 nm)

 

MCP-PMT Detectors

 HPPD-860 COOLEDMCP-PMT (R3809-50)

Typical IRF FWHM at 400 nm

40-45 ps 

40-45 ps

Shortest Measurable Lifetime 

40-45 ps 

5 ps

Wavelength Response

 220-860 nm

160-850 nm

Robust for Steady State Spectra

Yes

No

Amplifier + CFD

Integrated (no cable)

External

High Voltage Bias

Integrated (no cable)

External

Compatible with Phos and Steady State

Yes

No (requires second detector)

Temperature Control

Integrated TEC (air-cooled)

External (water-cooled)

Dark Count Rate (Cooled)

 < 200 cps 

 < 20 cps

PC Interface

N/A

N/A

 

NIR TCSPC Specifications

 

ModelSensorWavelength RangeTemporal ResponseDark CountCooling

NIR-R4

R5509-43 PMT

300 to 1,400 nm

1.5 ns (TTS)

< 25,000 cps

Liquid nitrogen

NIR-R7

R5509-73 PMT

300 to 1,700 nm

1.5 ns (TTS)

< 250,000 cps

Liquid nitrogen

NIR-H2

H10330-25 PMT

950 to 1,200 nm

400 ps (TTS)

< 2,500 cps

Thermo-electric

NIR-H4

H10330-45 PMT

950 to 1,400 nm

400 ps (TTS)

< 25,000 cps

Thermo-electric

NIR-H7

H10330-75 PMT

950 to 1,700 nm

400 ps (TTS)

< 250,000 cps

Thermo-electric

NIR-S1

Count-100N SPAD

400 to 1,000 nm

< 3 ns (TTS)

100 cps

None

 

FiPho TCSPC Electronics Specifications

Specifications

FiPho

FiPho-HR

Full Detectable TCSPC Lifetime Range

<20 ps to 30 sec

5 ps to 30 sec

TCSPC Converter Type

Digital TDC

Digital TDC and Analog TAC

TCSPC Bin Width

<15 ps

~ 250 fs

Phosphorescence Mode

MCS

MCS

Independent Stop Channels

1 to 4

1 to 4

Photon Streaming

Included

Included

FLIM Capable

Yes

Yes

 

Detailed FiPho Electronics Specifications

Specifications

FiPho (TDC, MCS)

FiPho-HR (TDC, TAC, MCS)

Full Detectable Lifetime Range

<20 ps to 30 sec

5 ps to 30 sec

TCSPC Time Range

<2 ns to 55 μs

<2 ns to 55 μs

Deadtime

5 ns

5 ns

TCSPC Bin Width

<15 ps

~ 250 fs

Electronics Jitter (FWHM)

30 ps

< 10 ps

TCSPC Histogram Size

Up to 16k

Up to 64k

Histogram Bin Depth

32 bit

32 bit

Independent Stops

1 to 4

1 to 4

Maximum Start Rate

100 MHz

100 MHz

Maximum Stop Rate

40 Mcps

40 Mcps

Operating Mode

Automatic Forward Timing

Automatic Forward or Reverse timing

Streaming Mode

Photon Streaming (Time-Tag)

Photon Streaming (Time-Tag)

MCS Bin Width

5 ns

5 ns

MCS Time Range

< 2.5 μs to 330 seconds

< 2.5 μs to 330 seconds

Maximum MCS Histogram Size

64k

64k

Acquisition and Analysis Macro Scripting

Yes

Yes

PC Interface

USB 3.0

USB 3.0

Software

EzTime, EzTime Image

EzTime, EzTime Image

HORIBA Scientific has a policy of continuous product development, and reserves the right to amend part numbers, descriptions and specifications without prior notice.

Dye‐protein binding monitored in a microliter volume using timeresolved fluorescence
Dye‐protein binding monitored in a microliter volume using timeresolved fluorescence
The potential health benefits stemming from the antioxidant activity of curcumin, commonly found in turmeric (Curcuma longa L), has attracted the interest of several research groups.
Stopped flow time‐resolved fluorescence study of serum albumin – curcuminoid binding
Stopped flow time‐resolved fluorescence study of serum albumin – curcuminoid binding
Rapid mixing accessories to perform stopped flow measurements have found application in characterizing interactions and reactions occurring in solution. Reactants are expelled from syringes, mixed and injected into a flowcell.
Fluorescence Anisotropy Studies
Fluorescence Anisotropy Studies
Polarized light striking a fluorescent molecule results in polarized fluorescence. This polarized emission gradually returns to unpolarized fluorescence depending on rotational diffusion and other factors. Anisotropy is directly related to the polarization, and is the ratio of the polarized light component to the total light intensity.
Measuring PL Upconversion Spectra and Lifetimes of Lanthanide-Doped Nanoparticles
Measuring PL Upconversion Spectra and Lifetimes of Lanthanide-Doped Nanoparticles
Upconverting lanthanide-based nanomaterials exhibit a unique fluorescence anti-Stokes shift, which enables them to convert NIR wavelength excitation into visible shorter wavelength emissions (NIR to UV-Vis).
Characterizing Lanthanides in Glasses for Optical Applications
Characterizing Lanthanides in Glasses for Optical Applications
Glasses are essential materials with a multitude of uses and many forms. In the area of optoelectronics there is an interest to modify the glass composition to favor the incorporation of lanthanide elements.
Upconversion of Lanthanide-containing glasses using DD‐980L excitation
Upconversion of Lanthanide-containing glasses using DD‐980L excitation
The phenomenon of upconversion is an optical process that takes in lower energy (longer wavelength) photons and emits higher energy (shorter wavelength) photons.
Measurement of carrier lifetime in perovskite for solar cell applications
Measurement of carrier lifetime in perovskite for solar cell applications
Hybrid perovskite photovoltaics (PV) show promise because of their good efficiencies, which can be around 20%. Along with their PV characteristics, perovskite materials exhibit a high degree of radiative recombination.
Monitoring Whole Leaf Fluorescence Using Time‐resolved Techniques
Monitoring Whole Leaf Fluorescence Using Time‐resolved Techniques
Light incident on a leaf can be absorbed by chlorophyll to commence the photosynthetic cycle. Excess energy can be liberated as heat or by emission of fluorescence and this can be used to assess the efficiency of the photosynthetic process.
The Measurement of Singlet Oxygen Lifetime Sensitized using Rose Bengal
The Measurement of Singlet Oxygen Lifetime Sensitized using Rose Bengal
The study of singlet oxygen (1O2) is of interest, principally, as it is a highly reactive species. It can be produced by photosensitisation, usually of a molecule such as a dye or porphyrin. Thus, by the appropriate selection of sensitiser, the presence of oxygen and light, 1O2 can be selectively generated. From a biological aspect it has the ability to damage and destroy cells, which has lead to interest in its use as an anticancer agent in photodynamic therapy (PDT).
Effect of temperature on HSA structure inferred using time-resolved room-temperature phosphorescence
Effect of temperature on HSA structure inferred using time-resolved room-temperature phosphorescence
To access intrinsic amino acids, such as tryptophan, as probes, the UV excitation wavelengths for pulsed phosphorescence measurements have long been the preserve of low-repetition-rate gas-filled lamps or larger laser systems. Recent developments have enabled the use of interchangeable semiconductor diodes...
Plasmon enhancement of protein fluorescence by silver nanostructures
Plasmon enhancement of protein fluorescence by silver nanostructures
The use of metal surfaces in conjunction with fluorescence molecules employing a plasmon effect, sometimes referred to as metal enhanced fluorescence, can be advantageous because of the possible enhancement of photophysical properties.
Investigating photocleavage using time‐resolved emission spectra
Investigating photocleavage using time‐resolved emission spectra
The choice of protecting group is of crucial importance in the success of many steps in organic synthesis and the manipulation of polyfunctional molecules, since they can prevent the formation of undesired side products and reactions.
Time‐resolved luminescence of security inks from the UV to NIR
Time‐resolved luminescence of security inks from the UV to NIR
The use of security features, such as luminescent inks, has increased significantly in an attempt to prevent fraud and counterfeiting of materials and goods.
Elucidating Local Viscosity Using Fluorescence Lifetime Measurements
Elucidating Local Viscosity Using Fluorescence Lifetime Measurements
Certain fluorescent molecules, known as molecular rotors, can be employed to estimate the local (nanoscale) viscosity in microheterogeneous systems by measurement of their fluorescence lifetime. This can be advantageous over the usual fluorescence anisotropy method, as the measurement is simpler and faster to perform. This is demonstrated using the HORIBA Scientific TemPro fluorescence lifetime system to monitor the gelation of silica produced using the sol‐gel technique.
MCS and Protein Phosphorescence
MCS and Protein Phosphorescence
Tryptophan phosphorescence within protein molecules is gaining attention as a probe of protein dynamics and structure. The tryptophan phosphorescence lifetime, τ, varies with the protein molecule’s local environment and conformation.

Product Variants

Request for Information

Do you have any questions or requests? Use this form to contact our specialists.

* These fields are mandatory.

Product accessories

Related products

DeltaPro
DeltaPro

TCSPC Lifetime Fluorometer

FluoroMax Plus
FluoroMax Plus

Steady State and Lifetime Benchtop Spectrofluorometer

SpectraLED
SpectraLED

LED Phosphorescence Light Sources

CFD-2G
CFD-2G

Amplifier-Discriminator

DAS6
DAS6

Decay Analysis Software

DeltaPro
DeltaPro

TCSPC Lifetime Fluorometer

Duetta
Duetta

Fluorescence and Absorbance Spectrometer

Fluorolog-QM
Fluorolog-QM

Modular Research Fluorometer for Lifetime and Steady State Measurements

FluoroMax Plus
FluoroMax Plus

Steady State and Lifetime Benchtop Spectrofluorometer

Hybrid Picosecond Photon Detector (HPPD) Series
Hybrid Picosecond Photon Detector (HPPD) Series

HORIBA’s latest development in TCSPC detector technology

Nanosizer
Nanosizer

For Single‐Walled Carbon Nanotube Excitation‐Emission Map Simulation and Analysis

PPD Photon Counting Detector
PPD Photon Counting Detector

Single photons detection with picosecond accuracy

PrintQuest
PrintQuest

Automated Fingerprint and Palmprint Identification - AFIS & APIS Systems

SpectraLED
SpectraLED

LED Phosphorescence Light Sources

Aqualog - Water Treatment Plant Analyzer
Aqualog - Water Treatment Plant Analyzer

Automated Organic Analysis & Early Warning Sentinel

Aqualog Datastream
Aqualog Datastream

For Instant Water Quality Reports

CFD-2G
CFD-2G

Amplifier-Discriminator

DAS6
DAS6

Decay Analysis Software

DeltaDiode
DeltaDiode

TCSPC Pulsed Sources

DeltaPro
DeltaPro

TCSPC Lifetime Fluorometer

Duetta
Duetta

Fluorescence and Absorbance Spectrometer

FLIMera
FLIMera

SPAD array imaging camera for dynamic FLIM studies at real time video rates

Fluorolog-QM
Fluorolog-QM

Modular Research Fluorometer for Lifetime and Steady State Measurements

FluoroMax Plus
FluoroMax Plus

Steady State and Lifetime Benchtop Spectrofluorometer

Hybrid Picosecond Photon Detector (HPPD) Series
Hybrid Picosecond Photon Detector (HPPD) Series

HORIBA’s latest development in TCSPC detector technology

LabRAM Odyssey
LabRAM Odyssey

Confocal Raman & High-Resolution Spectrometer

LabRAM Odyssey Semiconductor
LabRAM Odyssey Semiconductor

Photoluminescence and Raman Wafer Imaging

LabRAM Soleil
LabRAM Soleil

Raman Spectroscope - Automated Imaging Microscope

Modular Raman Microscope
Modular Raman Microscope

Flexible Raman System

Nanolog
Nanolog

Steady State and Lifetime Nanotechnology EEM Spectrofluorometer

Nanosizer
Nanosizer

For Single‐Walled Carbon Nanotube Excitation‐Emission Map Simulation and Analysis

PPD Photon Counting Detector
PPD Photon Counting Detector

Single photons detection with picosecond accuracy

QuantaPhi-2
QuantaPhi-2

PLQY Integrating Sphere

SpectraLED
SpectraLED

LED Phosphorescence Light Sources

Universal Fiber Coupling Solution
Universal Fiber Coupling Solution

Connect any of our steady state and hybrid fluorometers to virtually any upright or inverted microscope!

XploRA™ PLUS
XploRA™ PLUS

MicroRaman Spectrometer - Confocal Raman Microscope