Model | OCMA-550 |
---|---|
Product name | Oil content analyzer |
Measurement method | Solvent extraction - non-dispersive infrared absorption analysis method |
Measured objects | Substances extracted from sample water into solvent and having infrared absorption near a wavelength from 3.4 μm to 3.5 μm |
Measurement range | 0 mg/L to 200 mg/L |
Resolution | For mg/L 0 to 99.9: 0.1, 100 to 200: 1 For mg/g, mg/kg, mg/PC 0 to 9.99: 0.01, 10.0 to 99.9: 0.1, 100 to 200: 1 Abs.: 0 to 1.000 Abs.: 0.001 Abs. |
Repeatability | 0 mg/L to 9.9 mg/L: ±0.4 mg/L ±1 dig. 10.0 mg/L to 99.9 mg/L: ±2.0 mg/L ±1 dig. 100 mg/L to 200 mg/L: ±4 mg/L ±1 dig. * For standard liquids |
Display method | 3.5 inches, 320×240 dots Backlight Color graphic LCD |
Calibration method | Select each optionally zero calibration and span calibration |
Extraction solvent | S-316 |
Amount of extraction solvent required | Approx. 6.5 mL |
Extraction method | Using the extraction solvent, and extracted manually outside the product |
Ambient operating temperature | 0°C to 40°C (no condensation) |
Power supply | 100 V to 240 V AC ±10%, 50/60 Hz |
Power consumption | 100 V to 240 V AC, Approx. 60 VA |
External dimensions | 195 (H) × 253 (W) × 293 (D) mm |
Mass | Approx. 5 kg |
External output | Output to a USB memory stick |
Functions |
|